Keep Your Hydrafacial Device Glowing

Follow these quick steps to maintain peak performance.

After Every Treatment:

Perform a system rinse for 20 seconds and empty Gunkie Jar. Syndeo only: Ensure the device is connected to Wi-Fi for the latest upgrades

At the End of Every Day:

Perform a system rinse for 60 seconds

At Least Once a Week:

Clean the manifold (or after every 50 Hydrafacial treatments, whichever comes first). Clean the handpiece and the exterior of the device, including power entry and Ethernet ports, to maintain hygiene

Every 6 Months:

Replace the handpiece and tubing

Quick Tips & 
Best Practices

  • Use only original, non-diluted Hydrafacial products to ensure consistent glowing results and to prevent product related damage to your device.

  • When puncturing serum bottles for manifold placement, use the Hydrafacial puncture tool and swirl it to allow proper airflow.

    Per your purchase agreement, your device should remain on your premises. However, if you must transport your device between treatment rooms, please:

1. Remove the Gunkie jar, handpiece, tubing, LED accessories and all serum bottles from the manifold

2. Have more than one person involved to assist you with such transportation,

3. Lift the device from the bottom of the back; never from the tray

4. Try your best to ensure the device remains upright and secure from movement throughout transport. If your device must be transported in another position, safely lay the device on its back and ensure you have properly secured the device to prevent any movement during transport and to avoid any damage. Damage to devices during transport are not covered under warranty.

  • Supplies

    Stock Up on Products for your device

    Shop Now